Abstract

The aim of this study is to find optimal gantry, collimator, and couch angles for performing single isocenter, multiple target stereotactic radiosurgery (SIMT-SRS). Nineteen angle sets were tested across seven linear accelerators for radiation-isocenter coincidence and off-isocenter coincidence. The off-isocenter Winston-Lutz test was performed to evaluate the accuracy of isocenter alignment for each angle set, and optimal angle sets as well as maximum off-isocenter distance to target for each angle set was determined. The influence of simulated patient weight on off-iso Winston-Lutz test accuracy was also inspected. The SNC MultiMet-WL phantom and MultiMet-WL QA Software v2.1 were used for the direct measurement and analysis of the off-iso Winston-Lutz test (also referred to as Winston-Lutz-Gao test). A two-step method was developed to ensure precise initial placement of the target. Nineteen beams were delivered at 6X energy and 2×2cm field size to each of six targets on the MultiMet Cube with couch kicks at five cardinal angles (90°, 45°, 0°, 315°, and 270°). To reduce imaging uncertainty, only EPID was used in target alignment and test image acquisition. A total of 200Ibs (90.7kg) of weight was also used to mimic patient weight. All tests were performed on both the free table and the weighted table. For two new TrueBeam machines, the maximum offset was within the 1mm tolerance when the off-iso distance is less than 7cm. Two older VitalBeam machines exhibited unfavorable gantry, couch, and collimator (GCC) angle sets: Linac No. 3 at (0,90,0), (0,270,0) and Linac No. 4 at (0,45,45) and (0,90,0). The C-Series Linacs failed in the majority of GCC angle sets, with Linac No. 5 exhibiting a maximum offset of 1.53mm. Four of seven machines show a clear trend that offset increases with off-isocenter distance. Additionally, the IGRT table was less susceptible to the addition of simulated patient weight than the ExactCouch. Among the seven linear accelerators addressed, newer model machines such as the Varian TrueBeam were more precise than older models, especially in comparison to the C-Series Linacs. The newer machines are more suitable for delivering SIMT-SRS procedures in all GCC angle sets, and the results indicate that newer TrueBeams are capable of performing SIMT-SRS procedures at all angle sets for targets of off-iso distances up to 7cm. The trend that offset between the target center and radiation field center increases with off-iso distance, however, does not always hold true across machines. This may be comprised by the EPID's severe off-axis horn effect. Lastly, the IGRT couch was less susceptible to patient weight compared to ExactCouch in the off-isocenter Winston-Lutz test.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.