Abstract

Purpose The purpose of this study was to evaluate the ability of the Myocor Coapsys device to restore leaflet apposition and valve competency off-pump in a canine model of functional mitral regurgitation (MR). Description The Coapsys device was surgically implanted in 10 dogs after MR induction by rapid ventricular pacing. The Coapsys consists of anterior and posterior epicardial pads connected by a subvalvular chord. The annular head of the posterior pad was positioned at the annular level to draw the posterior leaflet and annulus toward the anterior leaflet. Final device size was selected when MR was minimized or eliminated as assessed by color flow Doppler echocardiography. Evaluation All implants were placed off-pump without atriotomy, and mean MR grade was reduced from 2.9 ± 0.7 to 0.6 ± 0.7 ( p < 0.001) acutely. No hemodynamic compromise was noted. Conclusions The Coapsys device consistently and significantly reduced or eliminated functional MR acutely. Further study will be required to assess the chronic stability of the repair in this animal model.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.