Abstract

Ocular ultrasound is an invaluable tool for the evaluation of the eye and orbit. However, the eye and orbit are potentially sensitive to the thermal and mechanical effects of ultrasound. When performing B-mode imaging, dedicated ocular settings should be used. If these settings are not available, limiting the acoustic output to Food and Drug Administration (FDA) recommended maximum levels is strongly advised. Especially important is the acoustic output in spectral (pulsed) and color Doppler modes, which can exceed the FDA's maximum recommended levels for the eye. Adjusting settings to decrease acoustic output and limiting the time of the examination should be done when performing a Doppler examination. The acoustic output of shear wave elastography is significantly higher than FDA guidelines for the eye and should be considered experimental.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.