Abstract

The frequency of ocular microtremor (OMT) is related to the functional status of the brain stem, and thus OMT may be useful in the diagnosis and management of brain stem disorders. The paper discusses the design of an OMT measurement system and reports quantitative specifications for three portable systems. All systems use a piezo-electric element as the transducer, which measures the displacement of the sclera during eye rotations. The systems differ in the manner in which the signal is recorded. All systems can detect eye movements corresponding to displacements of the sclera ranging from 12 to over 3000 nm. The frequency responses of all systems are flat (< 2 dB deviation from peak response) between 20 and 150 Hz. The phase response shows deviations (< pi) at the extremes of this range, but qualitative comparison of input and measured signals demonstrates that phase distortion is not excessive. Thus all systems are acceptable for clinical studies involving OMT.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.