Abstract

The occurrence of left ventricular (LV) anodal activation during pacing with modern multipolar cardiac resynchronization therapy (CRT) systems has never been reported. The aim of our study was to demonstrate, by means of electrocardiogram (ECG) analysis, the occurrence of simultaneous cathodal-anodal LV capture with quadripolar LV leads. We studied 10 first-time recipients of a CRT device equipped with a quadripolar LV lead. During follow-up, standard supine 12-lead ECGs were obtained in available cathode-to-anode LV pacing configurations with a pulse amplitude equal to twice the pacing threshold. The occurrence of simultaneous cathodal-anodal LV capture was defined as the presence of variations in electrocardiographic ventricular activation (EVA) when the distal tip (cathode)-to-device can (anode) pacing configuration was compared with the distal tip (cathode)-to-proximal ring (anode) configuration. In eight patients, we found differences in EVA when different LV sites were paced through the unipolar LV tip and unipolar LV ring configurations. In these patients, a difference in EVA was detected in 61.5% (59 of 96) of the ECG leads (marked difference in 31.3%, slight difference in 30.2%). Changes in EVA between unipolar tip-to-can and bipolar tip-to-ring pacing that were suggestive of cathodal-anodal LV capture were found in six patients. In these patients, a total of 30 (41.7%) ECG leads showed a difference in EVA (marked difference in 20.8%, slight difference in 20.8%). In our experience, additional anodal capture by the proximal LV ring during LV pacing is provable in most recipients of a resynchronization device equipped with a multipolar LV lead.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call