Abstract

Interventional radiology (IR) procedures tend to be complex, which delivers high radiation exposure to patient. In the present study, we measured the radiation exposure dose [Hp(3)] in the eye using a direct eye dosemeter placed next to the physician's eye during procedures. Physicians wore a direct eye dosemeter just lateral to eyes and an additional direct eye dosemeter outside the radiation protective eyeglasses close to their eyes. Additionally, a neck glass badge was worn at the neck. Although we found a positive correlation between the left neck glass badge dose [Hp(0.07)] and the left eye lens dose [Hp(3)], the value of R2 of the regression equation were 0.62 and 0.71 (outside and inside). We thought that the exact eye lens dose might not be estimated from the neck glass badge. In conclusion, a correct evaluation of the lens dose [Hp(3)] using the direct eye dosemeter is recommended for tachyarrhythmia physicians.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.