Abstract

BackgroundPositron emission tomography using the [18F] fluoride metabolite combined with computerized tomography (F-PET/CT) can be used to analyze the metabolic status of the periprosthetic bone after surgery for total hip arthroplasty.MethodsTo obtain normal PET referent values, 44 patients with 5 models of well-functioning hip prosthetic components were analyzed by F-PET/CT, radiography, and clinical score. Another group of patients having painful total hip arthroplasty, but whose radiography showed no conclusive signs of loosening, was analyzed by F-PET/CT scans before revision surgery.ResultsPreoperative median F-PET scores of the bone metabolic activity were 6.65 (3.3-9.0) for the painful stem group and 1.85 (1.2-3.9) (P < .01) for the referent group having the same stem model. At revision surgery, the stems in the painful group were assessed to be loose. At 2-year follow-up, the revised patients were all pain free.ConclusionF-PET/CT may be a new diagnostic tool for assessing occult loose stems that are not seen by radiography.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.