Abstract

Simple, selective and sensitive analytical devices are of a great importance for medical application. Herein, we developed highly selective immunosensor for electrochemical detection of C-reactive protein (CRP) in blood sample. Branched polyethylenimine functionalized with ferrocene residues (PEI-Fc) was the main element of the recognition layer, which allowed: (i) covalent binding of an antibody in its most favorable orientation and (ii) voltammetric detection of the C-reactive protein. Anchoring of PEI-Fc to the electrode surface through the electrodeposition process leads to the formation of thin, stable and reproducible layers, which is extremely important in the case of electrochemical immunosensing. The proposed analytical device is characterized by high selectivity and sensitivity and can be successfully used in the concentration range of CRP from 1 to 5·104 ng mL−1. The determined limit of detection was circa 0.5 and 2.5 ng mL−1 for voltammetric and impedance analysis, respectively. The developed analytical device has also been successfully applied for the analysis of CRP level in rat blood samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.