Abstract

Background: The aim of this clinical study was to compare the occlusal caries detection (OCD) performance of the intraoral scanners (IOSs) Trios 4 (TIO, 3Shape) and Emerald S (EME, Planmeca) and the Diagnocam (DIA, KaVo) with the established visual (WHO) examination (VIS, reference method). Methods: Between 08/2022 and 02/2023, 60 children (mean age 9.6 ± 2.5 years) were examined as part of their regular dental checkups. OCD was performed at the tooth level, separately for primary and permanent unrestored teeth. Furthermore, two thresholds were analyzed: sound versus overall caries (pooled data of enamel and dentin caries, TH1) and pooled data of sound and enamel caries versus dentin caries (TH2). Results: The best agreement with the reference method (reliability) in both dentitions was obtained for DIA (ĸ = 0.829/ĸ = 0.846; primary/permanent teeth), followed by EME (ĸ = 0.827/ĸ = 0.837) and TIO (ĸ = 0.714/ĸ = 0.680). Similar results were shown for the diagnostic quality (sensitivity, specificity and area under the curve of the receiver operating characteristic curve), with higher values for TH1 than for TH2. Both IOSs and the DIA showed worse results than the reference method VIS. Conclusions: Currently, IOS should be used as an additional caries detection tool, especially for visualization, and cannot be recommended as a basic tool for diagnosis or invasive/noninvasive therapy decisions in OCD.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call