Abstract

Although the Codman-Hakim programmable valve is popular, several problems arising from its design have been described. The authors report an additional cause of shunt obstruction in the system. A 6-year-old girl who had received a ventriculoperitoneal shunt with the Codman-Hakim programmable valve system presented with worsening consciousness. The valve proved hard to flush, and emergency revision of the valve was performed. Examination of the extracted valve revealed that the pressure control cam had migrated into the outlet of the valve, thus causing the obstruction. A crack in the plastic housing surrounding the cam suggesting a past impact to the system was also revealed. These factors should thus be kept in mind as potential sources of obstruction of the valve system, especially in patients susceptible to episodes of head impact.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.