Abstract

Two digital radiograph systems for stereotactic mammography, one using a lens to couple a Lanex Regular screen to a back-illuminated charge-coupled device (CCD) and one using a fiber-optic taper to couple a Min-R Regular-type screen to a front-illuminated CCD, were evaluated with respect to observer performance. A contrast-detail phantom was imaged in a variety of equivalent exposure conditions on both systems. Six observers viewed images on a video monitor and recorded which objects were detected. Performance (percent correct detections) with the lens-coupled system using the Lanex Regular screen was significantly higher than with the fiber-optic-coupled system using the Min-R Regular-type screen. Differences in absorption efficiencies of phosphors used, as well as differences in design of the two cameras, can explain differences in observer detection performance.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.