Abstract
In consecutive patients with suspected venous thromboembolism the interobserver variability of the SimpliRED D-dimer test was evaluated by two observers who independently scored one plate, the between assay variation was performed simultaneously by a third independent observer, who assessed a second plate. The biological variation was studied, 1-4 hours later by an independent evaluation. A total of 155 patients entered the study, venous thromboembolism was present in 42 patients (28%). The interobserver variability was 2/83 samples, with a kappa of 0.95 (95% confidence interval 0.88-1.0). The between assay variation was 2/98, with a kappa value of 0.96 (95% confidence interval 0.90-1.0). When testing the biological variation the observers disagreed in 2 of 69 patients (3%). The SimpliRED D-dimer assay has a good to excellent interobserver variability, between assay variation and reproducibility.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.