Abstract

The study focused on changes in elasticity that accompany water storage of a synthetic orthodontic elastomer. We plotted loading and unloading curves to permit the direct measurement of instantaneous elastic recovery (IER) and permanent set (PS) and the calculation of delayed recovery (DR). We obtained baseline data by testing dry as received material. Comparable tests were performed on material that had been stored in water at 37 degrees C for one, seven, 14, 42, and 70 days. Gas chromatography-mass spectrometry was used for analysis of organic substances leached from the elastomer by water. A two-way ANOVA revealed that extension distance and water storage duration affected load requirement, IER, PS, and DR. The presence of leachable organic substances in 14-, 42-, and 70-day storage water was evidence of time-dependent matrix decomposition. Findings from tests of elastic performance and analysis of specimen storage water indicate that exposure of the elastomer to water leads first to weakening of noncovalent forces and subsequently to degradation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.