Abstract
X-ray ptychography, a scanning coherent diffraction imaging method, was used to reconstruct images of a "Siemens star" test pattern with amplitude and phase contrast. While studying how the use of illumination with an increased bandwidth results in clear improvements in the quality of image reconstructions, we found that an artificial change in the overall distance scale factor of the algorithm leads to a systematic response in the image, which is reproduced with an incorrect number of spokes. This pathology is explained by the conflict between the length scales set by the scan and by the diffraction patterns on the detector.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.