Abstract

To assess patient radiation during catheter ablation procedures and operator differences. From 84 patients (51 males, age 63 ± 10 years) undergoing complex catheter ablation by three experienced operators we collected: body mass index (BMI), procedure type and time, fluoroscopy time, dose area product (DAP), air kerma and X-ray system setting (cine, collimation and angiographic imaging angle). A new factor, fluoroscopy DAP-fluoroscopy time ratio, was introduced to compare operator differences. The results show the average procedure time was 179 (± 57) minutes (min), fluoroscopy time was 31 (± 21) min, DAP was 26.4 (± 19.6) Gy.cm(2) and air kerma was 0.26 (± 0.19) Gy. Procedure types were: pulmonary vein isolation (PVI) (52 %), redo PVI (11 %), pulmonary vein ablation catheter (PVAC) (14 %), ventricular tachycardia (VT) (8 %) and others (15 %). Inter-operator difference was observed in fluoroscopy and cine usage. Fluoroscopy DAP-time ratios showed a similar level of patient radiation dose rate by operator A and B (correlation: 0.89), and a significantly higher dose rate by operator C (correlation: 0.20, p<0.001; 0.26, p<0.01, to operator A and B). In conclusion, operators should be aware of patient radiation exposure levels and the influencing factors. Inter- and intra-operator differences can be measured and bench marked for improvement in X-ray efficiency and patient radiation reduction.