Abstract

The influence on the effectiveness of sterilization by low-temperature plasma sterilizer CASP-80A was investigated so as to provide a theoretical basis for reducing medical costs and achieving ideal sterilization effectiveness. To conduct the on-site simulation test, a clinical material sterilization test and a test of the influence of organic substance were conducted, the former by using the representative of Bacillus Stearothermophilus, preparing the bacteria-contaminated carrier through polytetrafluoroethylene (PTFE) simulated hose endoscopes, and the latter by using calf serum as the influence factor of the organic substance. The results show that the CASP-80A low-temperature plasma sterilizer could achieve effective sterilization by either the short-cycle or the long-cycle sterilization method depending on different materials, apparatus, and extent of contamination. The organic substances could influence the effectiveness of sterilization by the low-temperature plasma (H2O2) sterilizer.