Abstract

Deuterium–tritium (DT) reaction rates of imploding capsules have historically been measured using neutron detectors. Temporal resolution is limited by the size of the detector and distance from the source to detector. The reaction rates can also be measured using the 16.7 MeV gamma ray, which is produced by the same DT reaction, but statistically far less often than the 14.1 MeV neutron. Cherenkov detectors detect gamma rays by converting the gamma rays to electrons, which in turn produce Cherenkov light and record this visible light using a fast optical detector. These detectors can be scaled to large volumes in order to increase detection efficiency with little degradation in time resolution, and placed well away from the source since gamma rays do not suffer velocity dispersion between the source and detector. Gas-based Cherenkov detectors can also discriminate against lower-energy photons produced in and around the target. A prototype gas Cherenkov detector has been built and tested for detector response at the Idaho State University electron linear accelerator (LINAC). Later tests at the University of Rochester’s OMEGA laser facility proved the feasibility of this diagnostic by successfully recording the 16.7 MeV gamma ray. A second version has now been built to demonstrate the temporal resolution.