Abstract
The scanning Kelvin probe force microscope (SKPM) is a member of the scanning probe microscope (SPM) family, and was derived from the non-contact atomic force microscope (NCAFM) technique. The contact potential difference (CPD) originating from the work function difference between the tip and sample surfaces can be measured using SKPM with simultaneous observation of the topography image. Using SKPM the surface of semiconductor device has been observed to measure the dopant concentration as the CPD difference in two dimensions. Most of the SKPM measurements are acquired in the atmosphere. The lateral resolution is achieved less than 100 nm.On the other hand, Scanning capacitance microscope (SCM) of contact mode is now most popular method to measure the dopant concentration in semiconductor device. However, the lateral resolution of SCM is limited, because of the fact that the contact mode cannot achieve a true atomic resolution, the fact that the capacitance is measured through the insulating layer on surface, and the fact that the tip and sample is damaged by the high electric field to be applied between the tip and sample. So SCM never achieves the atomic level resolution.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.