Abstract

Cochlear implant (CI) recipients often struggle to understand speech in reverberant environments. Speech enhancement algorithms could restore speech perception for CI listeners by removing reverberant artifacts from the CI stimulation pattern. Listening studies, either with cochlear-implant recipients or normal-hearing (NH) listeners using a CI acoustic model, provide a benchmark for speech intelligibility improvements conferred by the enhancement algorithm but are costly and time consuming. To reduce the associated costs during algorithm development, speech intelligibility could be estimated offline using objective intelligibility measures. Previous evaluations of objective measures that considered CIs primarily assessed the combined impact of noise and reverberation and employed highly accurate enhancement algorithms. To facilitate the development of enhancement algorithms, we evaluate twelve objective measures in reverberant-only conditions characterized by a gradual reduction of reverberant artifacts, simulating the performance of an enhancement algorithm during development. Measures are validated against the performance of NH listeners using a CI acoustic model. To enhance compatibility with reverberant CI-processed signals, measure performance was assessed after modifying the reference signal and spectral filterbank. Measures leveraging the speech-to-reverberant ratio, cepstral distance and, after modifying the reference or filterbank, envelope correlation are strong predictors of intelligibility for reverberant CI-processed speech.

Full Text
Published version (Free)