Abstract
to investigate discrepancies in ECD measurements and to define objective criteria for the evaluation of a donor cornea before full-thickness grafting. In a hundred cadaver eyes, discrepancies in ECD measurements by different methods (specular microscopy, confocal microscopy, and keratoanalyzer) were studied and objective evaluation criteria developed along with an optimal algorithm of pretransplantation assessment. Digital fluorescence microscopy was chosen as the reference method. It has been established that a triple measurement average obtained with any of the tested methods is informative enough as to the state of the donor cornea. The highest ECD values were obtained with specular microscopy, the lowest - with confocal microscopy. For reliable evaluation of donor corneas, we recommend that the average of a triple ECD measurement be taken using one of the mentioned methods.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.