Abstract

Functional residual capacity (FRC) is an important diagnostic parameter measured using analyzer. Since analyzer is expensive as well as cumbersome for use of noisy vacuum pump, the FRC measurement becomes possible only in large well-equipped hospitals. The present study introduced a new wash-out technique to measure FRC by analysis, which is relatively cheaper and much simpler to apply. Slower response was compensated for high frequency to be coincided with response, thereby enabled indirect, but accurate concentration measurement. FRC was estimated by continuous integration of expired volume obtained with air flow signal. Experiment with 3 L syringe, a standard calibration device recommended by the American Thoracic Society, demonstrated less than 1% error at 0, 1, and 2 L. Correlation coefficient was almost ideal, guaranteeing linear estimation of FRC. The present technique is inexpensive and simple to apply, thus should he of great convenience.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.