Abstract

An upgrade of the structure of magnetic resonance flowmeter–relaxometer is proposed to improve the metrological characteristics. The improved structure of the magnetic resonance flowmeter–relaxometer is based on the modulation of static magnetic field Н0 by ac magnetic field in the region of the nutation coil. The experiments show that the measurements of liquid flow q in the proposed regime make it possible to eliminate the effect of measurement errors related to variations in amplitude and phase of the NMR signal inverted by 180° under rapid variations in quantity q and increase the measurement dynamic range of quantity q by 30%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.