Abstract
Numerical solutions for the steady axisymmetric flow through a disk-type prosthetic heart valve were obtained for Reynolds numbers from 50 to 600. A nonuniform mesh in both directions was used and the finite difference equations in vorticity and stream function were solved explicitly. Stream function, vorticity, and shear and normal stress plots are presented. These detailed results clearly identify regions of very high shear and normal stresses, regions of very low or very high shear stress at the walls and the extent of separated or reverse flow regions. The length of the separated flow region downstream of the disk agreed very well with experimental data. The maximum value of the shear stress occurred on the upstream corner of the disk.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.