Abstract

This study presents an investigation of an innovative microfluidic flow separator using both numerical and experimental approaches to calibrate contrast-enhanced ultrasound scanners. Numerical simulations were conducted using Lagrangian particles tracking and passive scalar transport methodologies using the OpenFOAM software. The experimental validation confirmed the accuracy of the numerical simulations, particularly at an imposed total pressure of , showing an excellent agreement in particle distributions. The study emphasizes the computational efficiency and modeling of passive scalar transport, providing valuable understanding into the behavior of scalar quantities in microfluidic systems. An optimized diffusion coefficient value of was identified, showing its critical role in achieving accurate simulation results and optimizing the performance of microfluidic flow separators for contrast-enhanced ultrasound scanner calibration.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.