Abstract

This study presents a numerical investigation of the flow field in a screw pump designed to circulate biological fluid such as blood. A simplified channel flow model is used to allow analysis using a Cartesian set of coordinates. Finite analytic (FA) numerical simulation of the flow field inside the channel was performed to study the influence of Reynolds number and pressure gradient on velocity distribution and shear stresses across the channel cross-section. Simulation results were used to predict flow rates, circulatory flow and the shear stresses, which are known to be related to the level of red blood cell damage (hemolysis) caused by the pump. The study shows that high shear levels are confined to small regions within the channel cross-section, but the circulatory nature of the flow causes an increased percentage of blood elements to pass through the high shear regions, and increases the likelihood of cell damage.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.