Abstract

The present work proposes a novel MEMS capacitive accelerometer based microphone with enhanced sensitivity. The accelerometer design is simulated in COMSOL MULTIPHYSICS 4.2, and the complete model including the electronic detection system is implemented in SIMULINK. The accelerometer is proposed to serve as a microphone in the fully implantable hearing application which can be surgically implanted at the middle ear bone structure. Therefore, the design parameters such as size, weight and resonant frequency require deliberation. A novel analytical model considering the impact of the width of the microlever and the capacitive fringe effects on the performance of the sensor has been developed. The accelerometer is designed within a sensing area of 1 mm2 and accomplishes a nominal capacitance of 4.85 pF, an excellent capacitive sensitivity of 5.91 fF/g and voltage sensitivity of 11.2 mV/g.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.