Abstract
Two centrifugal cardio pumps with recirculation region and anti-recirculation region are designed by speed coefficient method. We can obtain the flow distribution of impeller and inside the shell at design conditions from geometric modeling and numerical simulation of two cardio pumps by using related software, and compared the interior shear stress distribution of two cardio pumps at the same condition. The results show that the maximum shear stress in the cardio pump with the recirculation region is lower about 10% than the pump with the anti-recirculation region and the region of the shear stress is exceed the critical shear stress (150Pa) which produced hemolytic is less about 2/3 than the cardio pump with the anti-recirculation region from the shear stress distribution. As a result a tiny destroys are caused to the blood cell and the pump with the anti-recirculation region model correspond with the demand of the haemophysiology, the research results provide the theory reference and basis in the optimization design for such artificial cardio pump.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.