Abstract

This article discusses the impedance method in the forward calculation in magnetic induction tomography (MIT). Magnetic field and eddy current distributions were obtained numerically for a sphere in the field of a coil and were compared with an analytical model. Additionally, numerical and experimental results for phase sensitivity in MIT were obtained and compared for a cylindrical object in a planar array of sensors. The results showed that the impedance method provides results that agree very well with reality in the frequency range from 100 kHz to 20 MHz and for low conductivity objects (10 S/m or less). This opens the possibility of using this numerical approach in image reconstruction in MIT.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.