We present results from the first stage of numerical modeling of implanted rotary blood pumps which can be used in pediatric heart surgery in patients undergoing the Fontan procedure. Two three-dimensional models of pumps − the centrifugal and axial types − were constructed. Head pressure-flow characteristics were obtained for each model and the effects of pump geometry on blood flow at an operating point of 2.5 L/min were evaluated. Stagnation zones were identified by quantitative assessment of the volume of fluid with flow rates of 0-0.1 m/s. The distribution of flow lines was used to identify vortex zones. Numerical modeling of fluid flow in pumps was run in Fluent ANSYS 19.0 computational fluid dynamics software.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.