Abstract

Strain gauge plethysmography (SGP) is a non-invasive method used in the detection of deep venous thrombosis (DVT). The technique is based on the measurement of calf volume changes in response to venous occlusion by a thigh cuff, the volume changes reflecting the rates of arterial inflow and venous outflow. A numerical model of the blood circulation within the limb and the response of this to a SGP test has been derived, based on treating the different parts of the circulatory system in the leg as resistance and capacitance elements. The simulation results were compared with clinical studies and support the ability of SGP to detect non-occlusive clots of more than 50-60% of the lumen, as well detecting calf vein occlusion. The non-linear behaviour of the venous compliance with intra-luminal pressure appears to be a particularly important factor within the model. In addition, increases in venous tone due to post-operative venospasm were shown to be a potential source of false positive results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.