Abstract
Integrating miniature pumps within microfluidic devices is crucial for advancing point-of-care diagnostics. Understanding the emergence of flow from novel integrated pumping systems is the first step in their successful implementation. A Purcell-like elasto-magnetic integrated microfluidic pump has been simulated in COMSOL Multiphysics and its performance has been investigated and evaluated. An elastic, cilia-like element contains an embedded magnet, which allows for actuation via a weak, uniaxial, sinusoidally oscillating, external magnetic field. Pumping performance is correlated against a number of variables, such as the frequency of the driving field and the proximity of the pump to the channel walls, in order to understand the emergence of the pumping behavior. Crucially, these simulations capture many of the trends observed experimentally and shed light on the key interactions. The proximity of the channel walls in the in-plane direction strongly determines the direction of net fluid flow. This characterization has important implications for the design and optimization of this pump in practical applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.