Abstract

Separation of microparticles and cells from a heterogeneous mixture is critical in the fields of medicine, engineering, biology, and biotechnology. In this paper, a novel design for continuous separation of three particles/cells in a two-component fluid flow by dielectrophoresis, as the most efficient separation method, is discussed. A new solver is developed in OpenFOAM to simulate the separation of three polystyrene particles, which are only slightly different in size (2 μm, 3 μm, and 3.5 μm), in a circular channel containing a two-component fluid flow. Simulation results showed that for mean velocity U0 = 200 μm/s, electric potentials >25 V are required to separate particles efficiently. Then, separation of three biological cells including red blood cells, white blood cells, and MDA-MB-231 breast cancer cells is investigated. To achieve simultaneously two goals of separating the cells and preventing the two fluids to be mixed, it was found that the effective electric potential should be within the range of 9 V to 12 V.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.