Abstract

Abstract Fractures or other major bone defects in the mandible are commonly treated by mounting a reconstruction plate at the fracture site. To avoid complications due to loosening of the implant or the fixation screws, electrical stimulation presents a possibility to accelerate bone healing. The aim of this study was to investigate different combinations of activated electrodes and plate designs regarding their feasibility for electrostimulation of the mandibular bone. The electric field distribution for multiple implant designs was computed using the finite element method. The results suggest that the electrode insulation renders a crucial parameter, which influences substantially the stimulation impact and its power consumption.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.