Abstract

IntroductionThere are data capture devices that attach to the FreeStyle Libre sensor and convert its communication from NFC (Near-field communication) to Bluetooth technology, generating real-time continuous glucose monitoring. The accuracy of hypoglycemia measurements displayed by smartphone apps using this device has not been established. MethodsStudy of diagnostic tests. Numerical accuracy was evaluated, utilizing the absolute difference with respect to capillary glucometry (ISO 15197:2015 standard) and clinical accuracy, using the Clarke and Parkes (Consensus) error grids, for glucose measurements less than 70mg/dL performed with the FreeStyle Libre system and with the digital estimation xDrip+ app, in diabetic patients managed with insulin therapy. ResultsTwenty-seven patients were included (TIR 73.4%, TBR70 5.6%), who contributed 83 hypoglycemic events. Numerical accuracy was adequate in similar proportions with the FreeStyle Libre system compared to the xDrip+ app (81.92% vs. 68.67%, p=0.0630). The clinical accuracy evaluation showed that 92.8% of the measurements for xDrip+ and 98.8% for FreeStyle libre met the criteria according to the Parkes (Consensus) grid (p=0.0535); and 79.5% and 91.6% of the measurements met the criteria according to the Clarke grid (p=0.0273), being higher with FreeStyle libre. ConclusionsThe use of the NFC-Bluetooth transmitter (Miao-Miao) associated with the xDrip+ app does not improve numerical or clinical accuracy for detecting hypoglycemic events in diabetic patients managed with insulin therapy, compared to the FreeStyle Libre device.