Abstract

A double J stent has been used widely these days for patients with a ureteral stenosis or with renal stones and lithotripsy. The stent has multiple side holes in the shaft, which supply detours for urine flow. Even though medical companies produce various forms of double J stents that have different numbers and positions of side holes in the stent, the function of side holes in fluid dynamics has not been studied well. Here, the flow rate and pattern around the side holes of a double J stent were evaluated in curved models of a stented ureter based on the human anatomy and straight models for comparison. The total flow rate was higher in the stent with a greater number of side holes. The inflow and outflow to the stent through the side holes in the curved ureter was more active than in the straight ureter, which means the flow through side holes exists even in the ureter without ureteral stenosis or occlusion and even in the straight ureter. When the diameter of the ureter changed, the in-stent flow rate in the ureter did not change and the extraluminal flow rate was higher in the ureter with a greater diameter.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.