Abstract

We investigate scintillation linearity of a commercial high pressure 4He gas detector using monoenergetic 2.8 MeV neutrons from a deuterium-deuterium fusion neutron generator. The scintillation response of the detector was measured for a range of recoil energies between 83 keV and 626 keV by tagging neutrons scattering into fixed angles with a far-side organic scintillator detector. Detailed Monte Carlo simulations were compared to experimental data to determine the linearity of the detector response by comparing the scaling of the energy deposits in the simulations to the detector output. In this analysis, a linear scintillation response corresponds to a consistent value for the scaling factor between simulated energy deposits and experimental data for several different scattering angles. We demonstrate that the detector can be used to detect fast neutron interactions down to 83 keV recoil energies and can be used to characterize low-energy neutron sources, one of its potential applications.