Abstract

The Fukushima nuclear power plant accident resulted in the discharge of radioactive particulate material into the atmosphere. Consequently, several hospitals in Japan have observed black spots on x-ray computed radiography (CR) images caused by particulate radioactive fallout. These black spots have no effect on human health. To reduce the influence of black spots on CR images, we need to erase latent images on imaging plates (IPs) immediately before clinical use and read the IPs soon after the x-ray examination. Alternatively, the contaminated felt of a cassette can be cleaned or exchanged, if possible.