We aimed to test a new model of self-expanding tracheal stent so that it might be made available for clinical use. Using direct laryngoscopy, we placed polyurethane-coated, nitinol stents into the middle third of the trachea in 25 New Zealand rabbits. After a mean observation period of 26 days, we evaluated stent migration, degree of expansion, attachment, adherence, formation of granulation tissue, presence of inflammatory infiltrate, parietal involvement, and epithelial lining. The results showed complete radial expansion, little adherence to the tracheal mucosa, and low tissue attachment, as well as high rates of granuloma formation and stent migration. This new model proved to be biocompatible and showed a behavior similar to that of other stents on the market.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.