Abstract

Breast density is one of the strongest known breast cancer risk factors. Breast density has primarily been measured by either categorical scores of the proportion of opacity on a screening mammogram or as a ratio of the delineated dense area to the total breast area on a mammogram. These methods are limited by their subjectivity, their lack of absolute reference standards, and lack of automation for clinical use. We present a novel automated method for measuring breast composition we call Breast Compositional Density measured using single X-ray absorptiometry techniques, or BD(SXA). BD(SXA) measures breast compositional density by comparing the opacity on the mammogram to two reference standards imaged with each breast. We present the mathematical derivation of BD(SXA), methods for correcting for radiographic nonuniformity in the mammogram, and in vitro measures of accuracy and repeatability. Using phantoms, BD(SXA) has been shown to have a long-term repeatability of better than 2% for a breast composition range of 0 to 100% fibroglandular density.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.