Abstract

Potentiometric thiocyanate-selective sensors based on the use of three synthesized di-, tetra-, and hexa-imidepyridine derivatives as novel anionic neutral ionophores in plasticized poly(vinyl chloride) (PVC) membranes are described. The sensors exhibit significantly enhanced response towards thiocyanate ions over the concentration range 5×10 −6 to 1.0×10 −2 M with a lower detection limit of 0.3 μg ml −1 and slopes ranging from −55.6 to −58.3 mV per decade. Fast and stable response, good reproducibility, long-term stability, applicability over a wide pH range (2–8) and high selectivity for SCN − ion in the presence of 18 common anions are demonstrated. The sensors are used for direct potentiometric measurements of thiocyanate ions over the concentration range 0.2–580 μg ml −1 and for monitoring sequential titration of some metal ions (e.g. Ag +, Tl +, Cu 2+, Pb 2+) in binary and ternary mixtures. Sequential binding of these metal ions with SCN − ensures share stepwise titration curves with consecutive end point breaks at the equivalent points. Recoveries of 98.5–99.1±0.3% are obtained for metal ion concentrations of 0.06–4 mg ml −1.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.