Abstract
In rotator cuff repair, strong and long-lasting suturing techniques that do not require additional implants are needed. This study examines the ultimate load to failure and the Young's modulus at the suture-tendon interface for a novel single-loop knot stitch and double-loop knot stitch. These values are compared to those of the modified Mason-Allen stitch. Twenty-four infraspinatus muscles with tendons were dissected from porcine shoulders (twelve Goettingen minipigs). The preparations were randomly allocated to three groups of eight samples. Load-to-failure testing of the single-loop knot stitch, the double-loop knot stitch and the mMAS were performed using a Zwick 1446 universal testing machine (Zwick-Roell AG, Ulm, Germany). The highest ultimate load to failure for the three techniques occurred with the double-loop knot stitch with a median value of 382.2 N (range 291.8-454.2 N). These values were significantly higher than those of the single-loop knot stitch, which had a median value of 259.5 N (range 139.6-366.3 N) and the modified Mason-Allen stitch, which had a median value of 309.3 N (range 84.55-382.9 N). The values of the single-loop knot stitch and the modified Mason-Allen stitch did not differ significantly. Regarding the Young's modulus, no significant differences were found between the double-loop knot stitch with a median value of 496.02 N/mm² (range 400.4-572.6 N/mm²) and the modified Mason-Allen stitch with 498.5 N/mm² (range 375.5-749.2 N/mm²) with respect to the stiffness of the suture-tendon complex. The median value for the Young's modulus of the single-loop knot stitch of 392.1 N/mm² (range 285.7-510.6 N/mm²) was significantly lower than those of the double-loop knot stitch and modified Mason-Allen stitch. This in vitro animal study demonstrated that both the single-loop knot stitch and the double-loop knot stitch have excellent ultimate load-to-failure properties when used for rotator cuff repair. The introduced single-loop knot stitch and double-loop knot stitch offer an alternative to other common used stitch techniques in rotator cuff repair.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.