Based on poly(vinyl chloride) membranes, a novel miniaturized screen-printed all-solid-state copper(II)-selective electrode has been developed for applications in environmental monitoring. Performance and applicability of the ion-selective electrode (ISE) have been proved by potentiometric investigations. Conducting polymers were used as intermediate layers and as solid contacts between the ion-selective membrane and the graphite transducer. The ion-complexing reagent 2-mercapto-benzoxazole was incorporated into poly(vinyl chloride) membranes. In the concentration range 10-6 - 10-2 mol/L, the ISE exhibited a linear Nernstian potential response to copper(II) with an average slope value of 28 mV/decade. The detection limit was 3 × 10-7 mol/L. The electrode exhibits a short response time ( electrode has also successfully been used in potentiometric, complexometric titrations with ethylenediaminetetraacetic acid.