Abstract

AbstractA novel electrode was fabricated for the quantitation of Fe2+ ion. It was based on the covalent attachment of the gemifloxacin molecule (the recognition element) to the surface of MWCNTs to be incorporated as an electroactive material. Linear response of Fe2+ ions was found in the concentration range of 1×10−2 mol L−1 to 1×10−8 mol L−1 with a Nernstian slope of 30.37 ±0.3 (mV/decade) and attained a stable response within 5 s. The sensor exhibited LOD value of 4.8×10−9 mol L−1. It was applied in the monitoring of Fe2+ concentration in multi‐vitamins tablets, tap water and milk samples with acceptable recovery ranged from 94.00 % to 102.00 %.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.