Abstract

A new dyed polyvinyl alcohol (PVA) film dosimeter based on 3-(4,5-Dimethylthiazol-2-yl)-2,5-diphenyltetrazolium bromide (MMT) tetrazolium dye is proposed in this study for measuring high gamma radiation dose. Gamma cell irradiator that contains Co-60 gamma-ray source was used to expose the novel MMT-PVA films to different doses up to 25 kGy. The changed in optical property of irradiated and unirradiated films were characterized by UV–Vis spectrophotometer. The results show that the dose sensitive and the linear range of irradiated films were increased considerably with increase of MMT concentration from 1 to 5 mM. The dose response of dyed PVA film changed substantially with changing relative humidity (12–74%) as well as irradiation temperature (10–40 °C). The absorbance of the un-irradiated films does not change up to 10 days in dark while a significant increase in their absorbance was reported for similar films under fluorescent light. The irradiated dosimeters that kept in dark showed a perfect stability for 54 days. It was found that no obvious impact of dose rate on the irradiated MMT-PVA film dosimeters.