Abstract

Continuous blood pressure (BP) monitoring is important for the prevention and early diagnosis of cardiovascular diseases. Cuffless BP estimation using pulse arrival time (PAT) via a mathematical model which enables continuous BP measurement has recently become a popular research topic. In this study, simultaneous biomedical signals from ten healthy subjects were acquired by electrocardiogram (ECG) and photoplethysmogram (PPG) sensors and the continuous reference BP data were collected by a cuff-based Finometer PRO BP monitor. A hierarchical model was applied to estimate the parameters of a nonlinear model which in turn is used to estimate systolic blood pressure (SBP) using PAT with few calibration measurements. The mean absolute difference (MAD) between the estimated SBP and reference SBP is 4.35±1.43 mmHg using the proposed hierarchical model with three calibration measurements and is 4.36±1.17 mmHg with a single calibration measurement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.