Abstract

We previously presented a cardiopulmonary bypass (CPB) method with blood delivery via femoral artery cannulation for pediatric aortic arch repair operations using the Radifocus Introducer sheath. However, the flow rate with the Radifocus Introducer sheath is limited by accessory parts with the same structure having a smaller inner diameter among different sizes, rather than the sheath body. Therefore, we further devised a combination of the JELCO IV catheter, an extension tube, and a three-way stopcock with a larger opening to obtain more flow rate keeping the CPB circuit pressure significantly lower than when using the Radifocus Introducer sheath successfully.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.