Abstract

Evolved vapor coulometry is a measurement technique that selectively detects water and is used to measure water content of materials. The basis of the measurement is the quantitative electrolysis of evaporated water entrained in a carrier gas stream. Although this measurement has a fundamental principle—based on Faraday’s law which directly relates electrolysis current to amount of substance electrolyzed—in practice it requires calibration. Commonly, reference materials of known water content are used, but the variety of these is limited, and they are not always available for suitable values, materials, with SI traceability, or with well-characterized uncertainty. In this paper, we report development of an alternative calibration approach using as a reference the water content of humid gas of defined dew point traceable to the SI via national humidity standards. The increased information available through this new type of calibration reveals a variation of the instrument performance across its range not visible using the conventional approach. The significance of this is discussed along with details of the calibration technique, example results, and an uncertainty evaluation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.