Abstract

Two novel procedures for flow injection gradient titration with the use of a single stock standard solution are proposed. In the multi-point single-line (MP-SL) method the calibration graph is constructed on the basis of a set of standard solutions, which are generated in a standard reservoir and subsequently injected into the titrant. According to the single-point multi-line (SP-ML) procedure the standard solution and a sample are injected into the titrant stream from four loops of different capacities, hence four calibration graphs are able to be constructed and the analytical result is calculated on the basis of a generalized slope of these graphs. Both approaches have been tested on the example of spectrophotometric acid–base titration of hydrochloric and acetic acids with using bromothymol blue and phenolphthalein as indicators, respectively, and sodium hydroxide as a titrant. Under optimized experimental conditions the analytical results of precision less than 1.8 and 2.5% (RSD) and of accuracy less than 3.0 and 5.4% (relative error (RE)) were obtained for MP-SL and SP-ML procedures, respectively, in ranges of 0.0031–0.0631 mol L −1 for samples of hydrochloric acid and of 0.1680–1.7600 mol L −1 for samples of acetic acid. The feasibility of both methods was illustrated by applying them to the total acidity determination in vinegar samples with precision lower than 0.5 and 2.9% (RSD) for MP-SL and SP-ML procedures, respectively.