Abstract

AbstractA novel amperometric sensor based on a rigid graphite‐epoxy composite of which composition is near to the percolation is reported. The electrochemical response of the novel transducer material was evaluated in terms of reproducibility of the fabrication process and reproducibility and repeatability of the analytical signal. The signal to noise ratio was improved. atomic force microscopy (AFM) technique was used to obtain qualitatively information. Amperometric detection of chlorine in water was carried on at a set potential of −250 mV vs. Ag/AgCl. The developed flow injection analysis (FIA) system responded linearly to chlorine concentration between 0.15 mg L−1 and 4 mg L−1with a sensitivity of −0.20 μA L mg−1. The proposed system was applied to real samples from swimming‐pool water. No significant difference was observed regarding the standard method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.