The SARS‑CoV‑2 virus is released from an infectious source (such as a sick person) and adsorbed on aerosols, which can form pathogenic microorganism aerosols, which can affect human health through airborne transmission. Efficient sampling and accurate detection of microorganisms in aerosols are the premise and basis for studying their properties and evaluating their hazard. In this study, we built a set of sub-micron aerosol detection platform, and carried out a simulation experiment on the SARS‑CoV‑2 aerosol in the air by wet-wall cyclone combined with immunomagnetic nanoparticle adsorption sampling and ddPCR. The feasibility of the system in aerosol detection was verified, and the influencing factors in the detection process were experimentally tested. As a result, the sampling efficiency was 29.77%, and extraction efficiency was 98.57%. The minimum detection limit per unit volume of aerosols was 250 copies (102 copies/mL, concentration factor 2.5).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.